Directions for use

Chapter 11 Specifications

191

Table 35. Setup window

Arrhythmia Alarms

Features

Customizable default

Factory default

Alarm level

Medium for Pause, off for all
other arrhythmia types except
VFib, asystole and V-Tach

Lethal for VFib, asystole, V-Tach

Arrhythmia alarm limits

Presets values

Presets values

V-Tach, 125; PVC run, 6; PVC/
min, 30; Tachycardia, 180;
Bradycardia, 45; Irregular Limit,
45

Print on alarm

Off for all

Options:
Irregular Limit
Single ECG (active with 5-wire cable
only)
Pacer Display
Analyze Pacers

Off

Off

Off

Off

PaceMaker NonCapture

Disabled

Leads used for arrhythmia analysis

Welch Allyn 1500 Patient Monitor,
Micropaq, and Propaq LT monitor 5-wire
cables:

Arrhythmia channel 0 uses lead II.

Arrhythmia channel 1 uses lead V.

Arrhythmia channel 2 uses lead III.

Propaq Encore, Propaq CS:
The leads that the user selects for ECG1
and ECG2 are used for arrhythmia
analysis.

Arrhythmia alarms for paced patients

Arrhythmia alarms depend on Acuity
System pacemaker detection vs. monitor
pacemaker detection.