Chapter 11 Specifications
Welch Allyn Acuity Central Monitoring System
Setup windows
Heart rate (HR)
For patients being monitored for arrhythmias,
the HR value is derived from the arrhythmia
library instead of from the patient monitor.
Table 33. Waveform Window
Features
Factory default
Waveform Window auto
open
In waveforms views:
1. Upon monitor connection to Acuity System if empty window
2. Upon alarm if open space is available OR replaces a non-alarming Waveform Window
Number of waveforms
displayed simultaneously
Up to two
Display priority of
waveforms
1. ECG
3. P2
5. P4
7. RESP
9. NIBP
2. P1
4. P3
6. CO
2
8. SpO
2
Number of vital sign
numerics displayed
Up to four. If four or more monitored at once, the third and fourth alternate in the same
zone; others appear if they alarm
Display priority of numerics
1. HR/PR
4. P2
7. CO
2
10. NIBP
2. SpO
2
5. P3
8. RR/BR
11. Temperature
3. P1
6. P4
9. IPI
Heart rate (HR)
For patients being monitored for arrhythmias, the HR value is derived from the arrhythmia
library instead of from the patient monitor.
Table 32. Virtual Monitor
Features
Customizable default
Factory default
Table 34. Setup window
—
Alarms
Features
Customizable default
Factory default
Display priority of vital signs
Up to three vital signs, in this order of priority:
1. Vital signs in alarm
2. Active vital signs (in the order listed in window)
Alarm limits
Presets values
Presets values
Micropaq monitor:
HR: 50 to 120 bpm
SpO
2
: 90% to 100%
For all monitors except the Micropaq monitor: Presets
are the factory default alarm limits of the patient
monitor, based on patient mode. For further information,
see monitor directions for use.
Stat Set values
Acuity Systems execute the Propaq Encore monitor
STAT SET function. For further information, see Propaq
Encore directions for use.
Auto Set values
Acuity Systems execute the Propaq Encore monitor
STAT SET function. For further information, see Propaq
Encore directions for use.