Chapter 11 Specifications

Welch Allyn Acuity Central Monitoring System

Table 42. Review window

Arrhythmia Events

Features

Factory default

Waveform display

1. Arrhythmia
2. ECG waveforms used at the time of arrhythmia analysis (one visible in Half Screen

view, more visible in Full Screen view depending on space and monitor type).

Period displayed

25 mm/s; period depends on sweep speed

The most recent seconds monitored appear at the right side of the window, or if another
review window was viewed first, the period is centered around the cursor bar position in
the first-viewed review window.

Presets values

ECG scale

1 mV

Sweep speed

25 mm/s

Blue cursor bar shows

Event types. Event markers are arrhythmia types, Arrhythmia Relearn and ST Relearn or
Learning Done and snapshot print.

ECG beat labels

Normal (

) and ventricular (

v

)

24-hour (or up to 96-hour)
time line indicates patient
and equipment state

Time line color matches icon color at all points in time

Number of deleted
arrhythmias allowed before
the Delete Selected button
becomes disabled

1000 per patient

Acuity System screen
views that support this
window

Full-screen and half-screen views

Printouts

Waveforms
Events