124

Chapter 6 Review or print patient data

Welch Allyn Acuity Central Monitoring System

Hemodynamic Report printouts

Hemodynamic Reports are a collection of calculations derived from the cardiac output
feature of the Welch Allyn 1500 Patient Monitor.

Print a Hemodynamic Report

To print a Hemodynamic Report

1.

Run a cardiac output measurement at the Welch Allyn 1500 Patient Monitor.

2.

Select Hemodynamic Calculations at the monitor.

3.

Select Transfer Data.

The monitor sends the Hemodynamic Report to the central station printer.

4.

If a re-print of the Hemodynamic Report is needed, place the cursor under printer icon
at the Waveform Review Window or the Graphical Trends Review Window.

Click Print and select Print Report.

A printout occurs at the Acuity System printer.

Review a Hemodynamic Report

A.

Patient information appears if the patient is confirmed in the Acuity System.

B.

Henodynamic calculations. This area displays the information generated by the Welch Allyn 1500
Patient Monitor regarding the hemodynamic calculations.

A

B