Directions for use
Chapter 9 Status icons
143
Patient Alarm
or
Telemetry Dropout
Take priority
over alerts.
Locations representing
monitors that are audibly
alarming in response to
one of these conditions:
•
Patient vital sign
limit violations
•
Patient monitor is
not communicating
with the Acuity
System.
[room #]
VT
ASYS
VF
APNEA
Limit vital sign
or
[room #] Dropout
Patient Alarm
Attend to the patient and
suspend or silence the
alarm tone. You can turn
off or reset alarm limits
also at the central station.
Then click the Silence/
Reset key at the patient
monitor. or Resume at the
central station Virtual
Monitor.
Telemetry Dropout
During this state,
Arrhythmia and ST alarms
are not indicated.
1. Click one of these
buttons:
•
SUSPEND to
suspend the alarm
tone for 90 seconds
•
RESET to stop the
alert and put the
monitor in the blue,
“patient intentionally
disconnected” state,
2. Bring the wireless
monitor within range
of a wireless access
point.
3. If the problem
persists, contact
qualified service
personnel.
No Continuous Heart/
Pulse Rate Source
Take priority
over alerts.
Locations representing
monitors that no longer
receive continuous heart/
pulse rate signals. If
monitoring began
without an HR/PR
source, the icon remains
blue.
[room #]
Source HR/PR
Reconnect the HR/PR
source at the patient
monitor.
Silence or suspend the
audible alarm tone.
Table 8. Icons and messages that represent patient and monitor status
Patient/monitor
status icon
Icon display
priority
Icon location on map
Messages in
hallway message
panels and at top
of Acuity Central
Station screen
Recommended action