Page 173

background image

Directions for use

Chapter 10 Troubleshooting and maintenance

169

Mobile Acuity LT Systems: Adjust localization parameters

If your Acuity System is a mobile (laptop computer) system, you can adjust these system
parameters.

Time zone

Acuity System language

Amended ID

Micropaq monitor language

Printout paper size

EN1060 compliance

Temperature units

Enable or disable the optional parallel printer

Set facility name (appears on printouts)

Set unit name (appears in upper right corner of Acuity system display)

Set printer ISO encoding

Set barcode variables

Set ECG filter

To adjust Mobile Acuity LT localization parameters

1.

Hold down the

key on your keyboard, move the cursor to the upper right corner

of the screen, and press the right mouse button.

2.

Click Quit Application, and then click Confirm.

3.

Wait for the login screen to appear.

The screen appears within one to two minutes.

4.

Type “iupdate”, and then press Enter.

5.

Follow the prompts to make the desired changes.

6.

Press q to quit.

7.

Take one of these steps:

If the system prompts you to reboot, take these steps:

a.

Choose 2 to quit.

b.

Reboot the system.

The system reboots and starts the Acuity application.

If the system prompts you to quit, take these steps:

a.

Choose 1 to quit.

Wait for the login prompt.

b.

Type “acuity”, and then press Enter.

You are logged in to the Acuity application.