4 Operation

766 IC Sample Processor

88

>final sequence
1 CTL:Rm: PUMP R/S 1
>changer settings
rack number 0
lift rate 1 12 mm/s
shift rate 20
>manual stop
CTL Rmt: **************
CTL RS232:
------------

Switch off 709 IC Pump

—— Settings for changer functions ——

—— Reaction to manual stop ———

Settings for 732 IC Detector

Setting of measurement parameters and creation of a program ac-

cording to section 3.3.2 of the 754 Instruction for Use. For the program

type, select

remote

instead of

cycle

.

Settings in the «IC Metrodata» program

Definition of an IC method without method associated remote control.

The analysis time

Duration

must be ca 1 min shorter than the analysis

time entered in the 732 IC Detector.

Procedure at program start

1. Enter parameter

number of samples

for 766 IC Sample Processor.

2. Create new queue in the program «IC Metrodata» and start it.
3. Start 766 IC Sample Processor with .