4.1 Fundamentals of operation
766 IC Sample Processor
39
Key
Normal state
Edit mode
Turn rack right
•
The <è> key turns the sample
rack one position to the right, i.e. in
the clockwise direction. The next
lowest beaker position is placed
under the needle.
•
The turning speed of the rack can
be defined in the Parameter menu.
•
The rotation of the rack can only be
carried out when the lift is at or
above the shift position.
Move the cursor one column
to the right
•
With the <è> key the cursor is
moved one column to the right in
an editing line with two parameters.
•
Any data which has been altered
will not be carried over during this
action (see
INSERT
Add a command line to a
sequence
•
Adds a new command line above
the current line in a sequence.
The
NOP
command (no operation)
automatically occupies this line and
has no function.
•
The lines following this line are
shifted one line downwards.
DELETE
Delete a command line in a
sequence
•
Deletes the current line in a
sequence.
•
The lines which follow shift upwards
by one line.
TOWER
SELECT
Select parameter
•
With the
data values can be selected for a
particular parameter in manual
operation.
•
With every repeated keystroke the
next value that can be selected is
displayed. The last value is followed
again by the first.
•
The data is accepted with