6 method "ancatseg – Metrohm 766 IC Sample Processor User Manual
Page 89

4 Operation
766 IC Sample Processor
84
Settings in the «IC Metrodata» program
1. Definition of anion method for channel 1 (analysis time, start pa-
rameters for IC instruments 732, 733 and 709, etc.), additional asso-
ciation of this method with the following time program for the 732/1
IC Detector:
Method / Equipment / Setup / Program / Text:
0.0
Puls_766_set
Set remote line 3 at 732/1 to 1 (s. below)
0.1
Puls_766_reset
Set remote line 3 at 732/1 to 0 (s. below)
0.2
Flag
end
Program end
Method / Equipment / Setup / Program / Remote configuration:
Puls_766_set
= **1*****
Set remote line 3 at 732/1 to 1
Puls_766_reset
= **0*****
Set remote line 3 at 732/1 to 0
Method / Equipment:
Start with method
2. Definition of cation method for channel 2 (analysis time, start pa-
rameters for IC instruments 732, 733 and 709, etc.). The analysis
time
Duration
must be shorter than the analysis time for the anion
method.
Procedure at program start
1. Create queue with cation method for channel 2 in the program «IC
Metrodata» and start it.
2. Start 766 IC Sample Processor with
3. Create queue with anion method for channel 1 in the program «IC
Metrodata» and start it.
4.6.6
Method "AnCatSeg"
Application
Simultaneous determination of anions and cations, additional insertion
of a defined air bubble between the samples.
Interconnection
see Fig. 14
Program of 766 IC Sample Processor
766 IC Sample Proc. 11111 5.766.0010
parameters
method AnCatSeg
number of samples: rack
>start sequence
1 CTL:Rm: INIT
>sample sequence
1 SCN:Rm : Wait1
2 SCN:Rm : Pump* ?
3 MOVE 1 : sample
4 PUMP 1.1 : 5 s
5 LIFT: 1 : work mm
6 CTL:Rm: FILL A 1
−
Report header with serial number and program ver-
sion
−
Method name
−
Number of samples to be processed (entire sample
rack)
−
Initialize remote interface
−
Waiting until 732 IC Det. sends signal on rem. line 3
−
Waiting until the two 709 IC Pumps run
−
Move needle to sample position
−
Aspirate air into the transfer tubing during 5 s
−
Place lift with needle to working position
−
Switch injection valve A at 733 to "Fill"