beautypg.com

Metrohm 766 IC Sample Processor User Manual

Page 22

background image

2.3 Attaching the accessories

766 IC Sample Processor

17

88

35

35

10

10

36

36

13

13

15

15

36

36

35

35

18

18

17

17

34

34

66

Fig. 9: Installing the pump tubing

66

PEEK compression fitting

(6.2744.010)

17

17

Snap-action lever

88

PEEK capillary (6.1831.050)

18

18

PEEK capillary (6.1831.060)

10

10

Pump tubing (6.1826.040)

34

34

PEEK compression fitting

(6.2744.010)

13

13

Tubing cartridge (6.2755.000)

35

35

Coupling (6.2744.030)

15

15

Contact pressure lever

36

36

Stopper (black-black)

3 Connection pump tubing – injection valve

At the 733 IC Separation Center, loosen the rotary nipple

screwed onto the interior side of connection 22

22 or 28

28.

Take PTFE suction tubing 84

84 (see Fig. 14 and Fig. 15 of the

732/733 Instructions for Use) completely out of connection 22

22

or 28

28 and unscrew from connection "1" of injection valve 66

66.

Pull the PEEK capillary 18

18 through the opening 22

22 or 28

28 of

the 733 IC Separation Center and screw onto connection "1"

of injection valve 66

66 using a 6.2744.010 PEEK compression

fitting.

Retighten rotary nipple on the interior side of connection 22

22

or 28

28 to fix the capillary 18

18.

4 Tubing connection injection valve – waste

Insert 6.2744.020 coupling (from 733 accessories) into con-

nection 21

21 or 27

27 of the 733 IC Separation Center.

Screw PTFE suction tubing 84

84 onto the 6.2744.020 coupling

attached to connection 21

21 or 27

27 and lead it into the waste

container.

In the case of the 733.0020 IC Separation Center with two injection

valves, it is possible to fill both sample loops from the same 766 IC

Sample Processor. For this, connection "1" of valve A (outlet of the

sample loop) must be connected to connection "2" of valve B (inlet of

the sample loop) using a 6.1803.040 PEEK capillary (15 cm).