6.5 Initialize data memory
766 IC Sample Processor
125
•
Prepare instrument for diagnostic test (see section 6.4.2).
•
If need be, press <9> key repeatedly until
diagnosis
>RAM initialization
•
Press
>RAM initialization
select: param
ü
>RAM initialization
select: config
ü
>RAM initialization
select: setup
ü
>RAM initialization
select: assembly
ü
>RAM initialization
select: all
ь
ь
ь
ь
ь
Set method parameters with default values.
Ã
Set config parameters with default values.
Ã
Set setup parameters with default values.
Ã
Set assembly parameters with default values.
Ã
Deletes all user-defined methods
Ã
•
Pressing the
dividual initialization alternatives are accessed using the
key, exit is with the
•
The table shows which parts of the data memory are affected by the
corresponding initialization alternatives. When the instrument reacts
with a system crash after being switched on (undefined display, no
reactions to keystroke, etc.) we recommend to carry out the initiali-
zation with the submenu
'all'
.
•
If need be, press the
>RAM initialization
select: all
•
Press
diagnosis
>RAM test
•
Press
The instrument quits the diagnostic menu and runs a power on reset.