3 Operating tutorial

766 IC Sample Processor

34

<ê>

>sample sequence
9 NOP

An empty line with a

'NOP'

-entry (no operation) always forms the end

of a sequence.

Exit the sample sequence with and go to the final se-

quence.

START

After all sample tubes have been processed, the final sequence is

executed.

>final sequence
1 CTL:Rm: PUMP R/S 1

In this line the pump drive at the 709 IC Pump is switched off. Press

.

START

>final sequence
2 CTL:Rm: PUMP 752 OFF

In this line the pump drive at the 752 Pump Unit is switched off.

Press .

QUIT

>final sequence
3 NOP

Now you have reached the end of the final sequence and have

completed the entire run of a sample series.

By pressing twice the 766 IC Sample Processor returns to

the normal state.

PARAM

7

SAMPLE

START

Now prepare some sample tubes and place them on the sample

rack. Prepare the IC system with 732 IC Detector, 733 IC Separation

Center, 709 IC Pump and 752 Pump Unit for the recording of chro-

matograms.

Enter the number of samples to be processed () and

define the position of the first sample (SAMPLE = 1).

Now you can start your first sample series with .