2.5 Connection of devices to the RS232 interface
766 IC Sample Processor
23
2.5
Connection of devices to the RS232
interface
2.5.1
General information on RS232 interface
Many different instruments may be connected via the serial RS232 in-
terface 32
32. In addition to all Metrohm instruments that support the
Metrohm remote control language (see section 5.2) any printer with se-
rial interface (or parallel interface and parallel/serial converter) or a per-
sonal computer (PC) may be connected. Any other measuring instru-
ment may be controlled via RS232 interface, as long as it supports
serial data transmission.
766
732
Fig. 16: Connection possibilities for the RS232 interface
In order to guarantee safe data transmission, it is important to set the
same RS232 interface parameters correctly for both instruments con-
nected (see section 2.5.2).
Control commands (examples)
CTL:RS &M;$G
starts a Metrohm instrument
CTL:RS &M;$S
stops a Metrohm instrument
PRINT: config
prints a configuration report to
a printer or PC
Scanning input data (example):
SCN:RS : *R"
waiting for readiness of a Metrohm
instrument
6.2125.060 cable
Printer cables, see section 2.5.2