Metrohm 766 IC Sample Processor User Manual
Page 58

4.2 Basic settings
766 IC Sample Processor
53
>rack definitions 2
type: M129-2
M129-2
…
…
Type description of the rack
and self-defined rack types (see section 4.5).
>rack definitions 2
work position 125 mm
0
…
125 mm
Working position of the needle
in mm from the upper stop.
Pressing
sition.
If you use the steel needle for 6.2743.050 sample tubes which are
sealed with 6.2743.060 PE caps, the working position must always be
set to 125 mm, since otherwise a vacuum can develop in the sample
tube and the sample will not be aspirated correctly.
>rack definitions 2
rinse position 125 mm
0
…
125 mm
Rinsing position of the needle
in mm from the upper stop.
Pressing
sition.
>rack definitions 2
shift position 0 mm
0 mm
Shifting position of the needle
in mm from the upper stop.
Only the value
'0'
is permissible.
>rack definitions 2
special position 0 mm
0
…
125 mm
Special position of the needle
in mm from the upper stop.
Pressing
sition.
>rack definitions 2
>>special positions
Submenu
Special positions submenu
>>special positions
special beaker 1 128
0
…
number of positions
Position of special beaker 1
>>special positions
special beaker 2 129
0
…
number of positions
Position of special beaker 2
etc. up to special beaker 8
Up to 8 special beaker positions can be defined
(see section 4.5).