6.4 Diagnosis

766 IC Sample Processor

123

6.4.9

Beeper

Prepare instrument for diagnosis (see section 6.4.2).

If necessary, press <ê> several times until

diagnosis

>beeper test

The beeper is turned on and off in an endless loop.

>beeper test

The test can be exited by pressing the or keys.

diagnosis
>rack code test

6.4.10 Rack code recognition

This diagnosis step checks the functionality of the sensors with regard

to their ability to automatically recognize the attached rack.

Note the code (magnets) on the underside of all the racks being

used. The following diagram shows the magnet fixture on the under-

side of a rack.

Magnets inserted

In this example
The code is:

000110

Magnet fixture as viewed from underneath

Pick up the rack and lay it on its side.

Prepare the instrument for diagnosis (see section 6.4.2).

If necessary press <ê> several times until

diagnosis
>power on reset

The sample changer runs through its start routine (initializing lift and rack positions). The ini-

tialization is important because it is vital for the following diagnosis step “rack code test“ that

the rack is in the basic position.

Prepare instrument for diagnosis (see section 6.4.2).

If necessary press <ê> several times until

diagnosis
>rack code test