Metrohm 766 IC Sample Processor User Manual
Page 81

4 Operation
766 IC Sample Processor
76
Magnets inserted
In this example
the code is:
000110
Rack bottom view
In order to assign a certain sample rack for different applications, cer-
tain properties or recognition data can be defined for up to 16 racks.
This is useful when a certain beaker size or the size of the sample se-
ries or a certain course of processing is to be predefined in an applica-
tion.
4.5.3
Rack data
The following recognition data can be defined for each rack:
Rack number
unique identification
Code
automatic rack recognition
Type
rack type / position table
Work position
working height of the needle
Rinse position
rinsing height of the needle
Shift position
turning height of the needle
Special position
additional height of the needle
Special beaker position reserved beaker positions
The Rack number serves to identify a rack. It can be chosen from 1 to
16. In a method a particular rack number can be assigned to the proc-
ess sequences (see section 4.3.2). This ensures that if the wrong sam-
ple rack is used, the automatic rack recognition will realize this and is-
sue a warning to the user.
The Code is used for automatic rack recognition. Make sure that this 6-
place binary code in the rack configuration agrees with the actual in-
serted magnet code on the rack. Rack codes can be changed at any
time. They must however, only be assigned to one rack. The assign-
ment of standard predefined codes of standard racks provided by
Metrohm should be avoided.