4 method "sp seg – Metrohm 766 IC Sample Processor User Manual
Page 87

4 Operation
766 IC Sample Processor
82
Settings for 732 IC Detector
Setting the measurement parameters.
Settings in the «IC Metrodata» program
Definition of an IC method without method associated remote control.
The analysis time
Duration
must be equal to the waiting time
WAIT
en-
tered under
sample sequence
in the 766 IC Sample Processor.
Procedure at program start
1. Create new queue in the program «IC Metrodata» and start it.
2. Enter parameter
number of samples
for 766 IC Sample Processor.
3. Start 766 IC Sample Processor with
4.6.4
Method "SP Seg"
Application
Processing a queue with the PC program «IC Metrodata» with the 766
IC Sample Processor as "Master", additional insertion of a defined air
bubble between the samples.
Interconnection
see Fig. 11 (without suppression) and Fig. 12 (with suppression)
Program of 766 IC Sample Processor
766 IC Sample Proc. 11111 5.766.0010
parameters
method SP Seg
number of samples: rack
>start sequence
1 CTL:Rm: INIT
2 CTL:Rm: PUMP 752 ON
>sample sequence
1 SCN:Rm : Pump1 ?
2 MOVE 1 : sample
3 PUMP 1.1 : 5 s
4 LIFT: 1 : work mm
5 CTL:Rm: FILL A 1
6 PUMP 1.1 : 120 s
7 CTL:Rm: ZERO 1
8 CTL:Rm: INJECT A 1
9 WAIT 1200 s
>final sequence
1 CTL:Rm: PUMP R/S 1
2 CTL:Rm: PUMP 752 OFF
>changer settings
rack number 0
lift rate 1 12 mm/s
shift rate 20
>manual stop
CTL Rmt: **************
CTL RS232:
------------
−
Report header with serial number and program ver-
sion
−
Method name
−
Number of samples to be processed (entire sample
rack)
−
Initialize remote interface
−
Switch on 752 Pump Unit for suppressor
−
Waiting until 709 IC Pump runs
−
Move needle to sample
−
Aspirate air into the transfer tubing during 5 s
−
Place lift with needle to working position
−
Switch injection valve A at 733 to "Fill"
−
Fill sample loop with sample during 120 s
−
Trigger autozero at 732 IC Detector
−
Switch injection valve A at 733 to "Inject"
−
Waiting time for recording and evaluation of the
chromatogram (must be adapted)
−
Switch off 709 IC Pump
−
Switch off 752 Pump Unit
−
—— Settings for changer functions ——
−
—— Reaction to manual stop ———