background image

4 Operation

766 IC Sample Processor

82

Settings for 732 IC Detector

Setting the measurement parameters.

Settings in the «IC Metrodata» program

Definition of an IC method without method associated remote control.

The analysis time

Duration

must be equal to the waiting time

WAIT

en-

tered under

sample sequence

in the 766 IC Sample Processor.

Procedure at program start

1. Create new queue in the program «IC Metrodata» and start it.
2. Enter parameter

number of samples

for 766 IC Sample Processor.

3. Start 766 IC Sample Processor with .

4.6.4

Method "SP Seg"

Application

Processing a queue with the PC program «IC Metrodata» with the 766

IC Sample Processor as "Master", additional insertion of a defined air

bubble between the samples.

Interconnection

see Fig. 11 (without suppression) and Fig. 12 (with suppression)

Program of 766 IC Sample Processor

766 IC Sample Proc. 11111 5.766.0010
parameters
method SP Seg
number of samples: rack
>start sequence
1 CTL:Rm: INIT
2 CTL:Rm: PUMP 752 ON
>sample sequence
1 SCN:Rm : Pump1 ?
2 MOVE 1 : sample
3 PUMP 1.1 : 5 s
4 LIFT: 1 : work mm
5 CTL:Rm: FILL A 1
6 PUMP 1.1 : 120 s
7 CTL:Rm: ZERO 1
8 CTL:Rm: INJECT A 1
9 WAIT 1200 s
>final sequence
1 CTL:Rm: PUMP R/S 1
2 CTL:Rm: PUMP 752 OFF
>changer settings
rack number 0
lift rate 1 12 mm/s
shift rate 20
>manual stop
CTL Rmt: **************
CTL RS232:
------------

Report header with serial number and program ver-
sion

Method name

Number of samples to be processed (entire sample
rack)

Initialize remote interface

Switch on 752 Pump Unit for suppressor

Waiting until 709 IC Pump runs

Move needle to sample

Aspirate air into the transfer tubing during 5 s

Place lift with needle to working position

Switch injection valve A at 733 to "Fill"

Fill sample loop with sample during 120 s

Trigger autozero at 732 IC Detector

Switch injection valve A at 733 to "Inject"

Waiting time for recording and evaluation of the
chromatogram (must be adapted)

Switch off 709 IC Pump

Switch off 752 Pump Unit

—— Settings for changer functions ——

—— Reaction to manual stop ———