4.6 Standard methods

766 IC Sample Processor

79

4.6

Standard methods

The following pages contain the listing of standard user methods in-

cluded with the instrument with explanations of the important com-

mands. A prerequisite for the use of these methods is a correct con-

figuration of the 766 IC Sample Processor for the 6.2041.430 sample

rack supplied (M129-2).
The user methods are designed for IC systems with the Metrohm in-

struments 732 IC Detector, 733 IC Separation Center, 709 IC Pump,

752 Pump Unit, 753 Suppressor Module, and 754 Dialysis Unit and the

PC evaluation program «IC Metrodata». The specific IC methods have

to be programmed at the 732 IC Detector or via PC program «IC Met-

rodata». Details for the connection of these instruments to the remote

connection 22

22 can be found in section 2.4.

It is recommended to work through every new method step-by-step

with the TRACE function, making adjustments where necessary, before

starting the method for the first time.

4.6.1

Method "PC"

Application

Processing a queue with the PC program «IC Metrodata» with the PC

as "Master".

Interconnection

see Fig. 11 (without suppression) and Fig. 13 (with suppression)

Program of 766 IC Sample Processor

766 IC Sample Proc. 11111 5.766.0010
parameters
method PC
number of samples: rack
>start sequence
1 CTL:Rm: INIT
>sample sequence
1 SCN:Rm : Wait1
2 SCN:Rm : Pump1 ?
3 MOVE 1 : sample
4 LIFT: 1 : work mm
5 CTL:Rm: FILL A 1
6 PUMP 1.1 : 120 s
7 CTL:Rm: ZERO 1
8 CTL:Rm: INJECT A 1
>final sequence
>changer settings
rack number 0
lift rate 1 12 mm/s
shift rate 20
>manual stop
CTL Rmt: **************
CTL RS232:
------------

Report header with serial number and program ver-
sion

Method name

Number of samples to be processed (entire sample
rack)

Initialize remote interface

Waiting until 732 IC Det. sends signal on rem. line 3

Waiting until 709 IC Pump runs

Move needle to sample

Place lift with needle to working position

Switch injection valve A at 733 to "Fill"

Fill sample loop with sample during 120 s

Trigger autozero at 732 IC Detector

Switch injection valve A at 733 to "Inject"

—— Settings for changer functions ——

—— Reaction to manual stop ———