6 Appendix

766 IC Sample Processor

120

The test can be exited with the or keys.

diagnosis
>display contrast test

After pressing the key the following display appears whereby the contrast of the

display continuously varies between bright and dark.

>display contrast test
766 IC Sample Processor

The test can be exited with the or key.

diagnosis

>key test

6.4.5

Keyboard

This diagnosis step allows all the keys on the keyboard to be tested for

their functionality.

Prepare instrument for diagnosis (see section 6.4.2).

If necessary, press <ê> several times until

diagnosis
>key test

>key test

Activate all the keys one after the other and check their reaction on

the display.

The corresponding matrix code appears in the display and a description of the main function of

the key pressed (for example, the following display should appear when the key is

pressed).

>key test
code 4 CONFIG

The test is exited by pressing the key twice.

diagnosis