6.1 Error messages
766 IC Sample Processor
111
6
Appendix
6.1
Error messages
When an error occurs the execution of the active command is inter-
rupted and an error message is displayed (display blinks). This must be
acknowledged with the
If an error occurs during processing of a sample series, the 766 IC
Sample Processor will then be switched into the "HOLD" state. After the
cause of the error has been rectified, the sample series can be contin-
ued with the next command in the active sequence by pressing the
can also be halted with
The list of possible error messages and their causes:
Error message
Cause
* battery low
The battery for the permanent storage
of user methods must be replaced.
* wrong rack
The rack positioned is not the one
that was assigned to the method un-
der
'parameters'
.
* rack data missing
No sample rack is in position or no
rack data can be found for the sam-
ple rack that is in place.
* raise lift first
Turning of a rack could not be carried
out because the lift was below the
defined shift position.
* user memory full
The memory for the user-defined
methods is full. Before saving a new
method, methods that are not used or
used only rarely must be deleted.