4 Operation

766 IC Sample Processor

86

4.6.7

Method "Preconc"

Application

Preconcentration of samples with very low concentrations in an pre-

concentration column (trace analysis).

Interconnection

see Fig. 13

Installation of the preconcentration column

The preconcentration column, consisting of the 6.1006.200 Metrosep

Anion preconcentration cartridge and the 6.2828.010 glass cartridge

holder, is installed at the injection valve of the 733 IC Separation Center

instead of the sample loop (see Fig. 18). It is important to install the

sample inlet from the 766 IC Sample Processor at connection 2 of the

injection valve in order to run the preconcentration column for the en-

richment and the sample injection in the opposite direction. If a larger

sample throughput is desired, install the optionally available 6.1826.040

pump tubing (inner diameter 1.6 mm, flow ca. 3 mL/min) at the peristal-

tic pump of the 766 IC Sample Processor.

1

2

4

5

3

6

1

2

4

5

3

6

Fig. 18: Installation of the preconcentration column

Program of 766 IC Sample Processor

766 IC Sample Proc. 11111 5.766.0010
parameters
method Preconc
number of samples: rack
>start sequence
1 CTL:Rm: INIT
2 CTL:Rm: INJECT A 1
>sample sequence
1 SCN:Rm : Wait1
2 SCN:Rm : Pump1 ?
3 MOVE 1 : sample
4 LIFT: 1 : work mm
5 PUMP 1.1 : ON s
6 WAIT 60 s
7 CTL:Rm: FILL A 1
8 WAIT 120 s

9 PUMP 1.1 : OFF s
10 CTL:Rm: ZERO 1
11 CTL:Rm: INJECT A 1

Report header with serial number and program ver-
sion

Method name

Number of samples to be processed (entire sample
rack)

Initialize remote interface

Switch injection valve A at 733 to "Inject"

Waiting until 732 IC Det. sends signal on rem. line 3

Waiting until 709 IC Pump runs

Move needle to sample position

Place lift with needle to working position

Switch on pump at 766

Waiting time 60 s

Switch injection valve A at 733 to "Fill"

Sample preconcentration during 120 s
(preconcentration time must be adapted)

Switch off pump at 766

Trigger autozero at 732 IC Detector

Switch injection valve A at 733 to "Inject"

Waste

Sample

Sample

Waste

Sample injection

Preconcentration

IC column

Preconc. column

Eluent

IC column

Eluent