beautypg.com

Metrohm 766 IC Sample Processor User Manual

Page 121

background image

6 Appendix

766 IC Sample Processor

116

6.3.3

Replacing the pump tubing

Pump tubings are consumable material with a limited lifetime and

should be replaced at regular intervals (approx. every 2 weeks under

continuous use).
The working life of pump tubing depends to a considerable extent on

the contact pressure. This is why the contact pressure must be cor-

rectly set as described in section 2.3.8. If the pump is to remain

switched off for a lengthy period of time the tubing cartridges should be

raised completely by loosening snap-action lever 17

17 on the right-hand

side (the pre-set contact pressure remains unchanged).
To replace a pump tubing proceed as follows:

1 Remove old pump tubing

Press contact pressure lever 15

15 on the tubing cartridge down

as far as it will go.

Release tubing cartridge 13

13 from holding clamp 16

16 by

pressing down snap-action lever 17

17 and remove from

mounting pin 12

12 at the 766 IC Sample Processor (see Fig. 1).

Remove old pump tubing.

2 Insert new pump tubing

Insert the new pump tubing 10

10 (6.1826.040) in the tubing

cartridge as shown in Fig. 6. The stopper 36

36 must click into

the corresponding holder on the left-hand side of the tubing

cartridge.

Place the tubing cartridge on mounting pin 12

12 and press

down on the right-hand side until snap-action lever 17

17 clicks

into position on holding clamp 16

16. Take care that no kinks are

formed in the pump tubing.

3 Set contact pressure

Press contact pressure lever 15

15 upwards until the solution

just starts to be drawn in. Then press contact pressure lever

upwards until it clicks once more to obtain optimal contact

pressure.

The 6.1826.0X0 pump tubing is made of PVC and must not be used

for rinsing with solutions which contain acetone. In such cases, rinse

with different pump tubing or a different pump.