Metrohm 766 IC Sample Processor User Manual
Page 20

2.3 Attaching the accessories
766 IC Sample Processor
15
27
27
33
33
32
32
22
Fig. 8: Adjusting the sample rack
2 Move to adjusting position
Press
'37'
and confirm with
22 is above the adjusting position 32
32 (position 37: first opening of
the middle row).
3 Check needle position
•
Press <
ê
> until the needle 22 is ca. 1 cm above the upper
level of the sample rack 27
27.
•
Check needle position: If the needle 2 cannot be lowered
unhindered through the upper hole of the adjusting position
32, continue directly with point 4.
•
Continue lowering the needle by pressing <
ê
> until the
needle 22 is ca. 1 cm above the lower level of the sample rack
27
27.
•
Check needle position: If the needle 2 cannot be lowered
unhindered through the lower hole of the adjusting position
32, continue directly with point 4.
•
Lower needle completely by pressing <
ê
>.
•
Check needle position: If the needle 22 is in the middle of the
lower hole, the sample rack must not be adjusted (continue in
this case with point 5).
4 Adjust sample rack
•
Loosen the three adjusting screws 33
33 on the lower level of the
sample rack using the 6.2621.100 Allen key
•
Carefully turn the two upper levels of the sample rack 27
27 by
hand until the lowered needle 22 is exactly in the middle of the
lower hole at the adjusting position 32
32.
•
Tighten the adjusting screws 33
33.
5 Move to rest position
Press
22 Steel neddle
(6.2624.000) or PEEK
needle (6.1835.000)
27
27 Sample rack
(6.2041.430)
32
32 Adjusting position 37
33
33 Adjusting screw