GX-2012 Operator’s Manual

Start Up, Leak Check Mode • 87

If calibration is not due or CL EXPRD is set to NO EFFECT, then
the following screen appears for a few seconds indicating when the
next calibration is due.

7. If BP RMNDR is turned on (factory setting is OFF), the screen that

appears next depends on how BP EXPRD is set in the Maintenance Mode
Menu (page 133).

If the unit is due for a bump test and BP EXPRD is set to CONFIRM
(factory setting), then the following screen displays.

The alarm LED’s and buzzer will pulse several times. After this,
press the RESET SILENCE button until you hear a beep to continue.

If the unit is due for a bump test and BP EXPRD is set to NOT USE,
then the following screen displays.

The GX-2012 cannot be used until a bump test has been performed
by selecting BUMP in the Maintenance Mode menu. See
“Performing a Bump Test” on page 111 for bump testing instructions.