Turning off the gx-2012, leak check mode – RKI Instruments GX-2012 Manual User Manual
Page 107
GX-2012 Operator’s Manual
Leak Testing • 99
The gas readings that were taken during the snap log are displayed and the
Station ID that was in use during the snap log is displayed along the
bottom of the screen. Snap log data that was taken in Leak Check Mode
operation will have 2 sequential screens that are part of one data set. One
screen is for the base reading and the other screen is for the peak reading.
The peak reading value will have PEAK displayed below the peak bar. If a
base data point appears without a peak data point after it, it is because a
peak data point was not taken. You can use the AIRS and (SHIFT)T
buttons to scroll through the different snap logs. The gas readings and
station ID will change but the snap log ID is not visible from this screen.
6. To exit snap log data viewing and return to Leak Check Mode Operation,
press and release the DISPLAY (ADJ) button.
Turning Off the GX-2012, Leak Check Mode
1. Press and hold the POWER ENTER button.
2. The buzzer will pulse for about three seconds and TURN OFF will appear
at the bottom of the screen.
3. Release the button when TURN OFF disappears.