38 • Measuring Mode, Alarms

GX-2012 Operator’s Manual

Resetting and Silencing Alarms

You can set the GX-2012’s gas alarms as latching or self-resetting alarms (see
“Updating the Alarm Latching Setting” on page 117) and the buzzer
operation as silenceable or not silenceable (see “Updating the Alarm Silence
Setting” on page 118).
• Self-resetting alarms (LATCHING set to OFF)

Self-resetting alarms automatically shut off and reset when the gas
reading falls below (or rises above for an oxygen low alarm) the alarm
setting. You cannot silence or reset self-resetting alarms.

• Latching alarms (LATCHING set to ON)

You can set latching alarms with or without Alarm Silence (see “Updating
the Alarm Silence Setting” on page 118).

With ALRM SLNC On and LATCHING On:

When the GX-2012 goes into a gas alarm, press the RESET SILENCE button
to silence the buzzer. If the gas concentration was still above the alarm level
when the button was pressed, the LEDs continue to flash, and the GX-2012

Sensor Failure*

FAIL appears at the

bottom of the display

and the failed

sensor(s) are

indicated

• Alarm LED arrays

flash

Buzzer sounds a double
pulsing tone (two pulses

in quick succession)

Clock Failure*

• FAIL CLOCK appears

on the display

• Alarm LED arrays

flash

Buzzer sounds a double
pulsing tone (two pulses

in quick succession)

System Failure*

• FAIL SYSTEM

appears on the

display

• Alarm LED arrays

flash

• Buzzer sounds a

double pulsing tone

(two pulses in quick

succession)

• A failure code appears

Microprocessor

Failure*

• NOTE: The unit will

not operate if this

alarm occurs.

• Heart indicator is

steadily on or not on

at all

• None

*

This alarm can also occur in Bar Hole Mode

Table 4: Alarm Types and Indications

Alarm Type

Visual Indications

Other Indications