GX-2012 Operator’s Manual

Replacing the LEL and CO Filters • 153

4. Insert the battery pack into the charging station as shown in Figure 25

below so that the metal contacts on the back of the battery pack come into
contact with the metal contacts in the charging station. When proper
contact has been made, the red LED on the charging station will turn on.

Figure 25: Putting the Battery Pack into the Charging Station

5. When a full charge has been reached, approximately 4 hours, the red

LED on the charging station will turn off. Remove the battery pack from
the charging station and unplug the charging station’s power cord from
the AC outlet.

Replacing the LEL and CO Filters

1. Verify that the GX-2012 is off.
2. Use a screwdriver to remove the three screws holding the belt clip to the

back side of the instrument. Two of these screws also secure the flow
chamber.

3. Remove the belt clip and unscrew the remaining screw holding in the

flow chamber.

4. Grab the exhaust tube and pull it gently away from the unit to slide the

flow chamber out of the unit. The CO (charcoal) filter may come out.

5. To replace the LEL filter, remove the old one from its recess in the sensor

gasket and replace it with the new one making sure that it is properly
seated. You do not have to remove the sensor gasket to replace the LEL
filter. It sits in a recess on top of it.

Battery Pack

SDM-2012

BUMP

GX-2012 OFF

COPY

POWER

CAL.

EDIT

ENTER

1SEC ON

3SEC OFF

CHARGE