GX-2012 Operator’s Manual
Replacing the LEL and CO Filters • 153
4. Insert the battery pack into the charging station as shown in Figure 25
below so that the metal contacts on the back of the battery pack come into
contact with the metal contacts in the charging station. When proper
contact has been made, the red LED on the charging station will turn on.
Figure 25: Putting the Battery Pack into the Charging Station
5. When a full charge has been reached, approximately 4 hours, the red
LED on the charging station will turn off. Remove the battery pack from
the charging station and unplug the charging station’s power cord from
the AC outlet.
Replacing the LEL and CO Filters
1. Verify that the GX-2012 is off.
2. Use a screwdriver to remove the three screws holding the belt clip to the
back side of the instrument. Two of these screws also secure the flow
chamber.
3. Remove the belt clip and unscrew the remaining screw holding in the
flow chamber.
4. Grab the exhaust tube and pull it gently away from the unit to slide the
flow chamber out of the unit. The CO (charcoal) filter may come out.
5. To replace the LEL filter, remove the old one from its recess in the sensor
gasket and replace it with the new one making sure that it is properly
seated. You do not have to remove the sensor gasket to replace the LEL
filter. It sits in a recess on top of it.
Battery Pack
SDM-2012
BUMP
GX-2012 OFF
COPY
POWER
CAL.