beautypg.com

Fail – RKI Instruments GX-2012 Manual User Manual

Page 71

background image

GX-2012 Operator’s Manual

Calibration Mode • 63

If any of the sensors cannot calibrate to the proper value, FAIL displays
along the bottom of the screen and the GX-2012 lists the sensor(s) that
failed to calibrate. In the example below, the OXY and H

2

S channels

failed calibration. The other sensors calibrated normally.

The buzzer and alarm lights activate. Press and release the RESET
SILENCE button after the buzzer and lights stop to reset the alarm and
return to the calibration menu. Attempt to calibrate again. If the failure
continues, investigate the cause. See “Troubleshooting” on page 143.

16. Disconnect the tubing from the probe.
17. Unscrew the demand flow regulator from the calibration cylinder.
18. If you have a 5 sensor unit and wish to calibrate the %volume

combustible gas sensor, go back to step 1 and repeat the process selecting
the %volume calibration gas value screen in step 12 and using a
%volume methane cylinder.
If you are finished calibrating, use the (SHIFT)T button to navigate to
the NORMAL menu item, then press and release the POWER ENTER
button to return to Measuring Mode.

Calibrating with the Single Calibration Method

This section describes calibration using the Single Calibration method.
To calibrate using the Auto Calibration method, see “Calibrating with the
Auto Calibration Method” on page 60.
The procedure below describes a span adjustment of one channel using a
calibration kit that includes a 4-gas calibration cylinder, a 0.5 LPM (liters per
minute) regulator, and non-absorbent sample tubing. The standard 4-gas
cylinder consists of 50% LEL methane, 12% oxygen, 50 ppm CO, 25 ppm
H

2

S, and a balance of nitrogen. In the example below, the CH

4

channel’s

span is adjusted. See “Parts List” on page 158 for available calibration kits
and calibration kit spare parts.
Calibration cylinders that contain only one target gas may also be used for
single calibration. For channels other than oxygen, RKI Instruments, Inc.
recommends that you use a cylinder that contains a concentration of 10-50%




%

ppm

MAINTENANCE

O

2

FAIL

H

2

S