94 • Leak Testing

GX-2012 Operator’s Manual

Peak Hold Mode

Peak Hold Mode allows the user to view the peak methane reading for the
operating session while viewing the current methane reading.
To enter Peak Hold Mode:
1. Turn the GX-2012 on as described in “Turning On the GX-2012, Leak

Check Mode” on page 84. Select the desired display range. In the example
below, 500 ppm has been selected.

2. Press and hold the RESET SILENCE button until the Peak Hold Mode

screen displays.

The current methane reading is displayed on the first line. The “PH”
shown on the second line indicates that the GX-2012 is in Peak Hold
Mode. The peak methane reading is displayed on the third line. The full
scale value is displayed on the fourth line.

3. The peak reading displayed while in Leak Check Mode will appear in the

Peak Hold Mode screen.

4. If the highest reading recorded is higher than the full scale value, the full

scale value will be displayed. If the full scale value is changed to
something higher, the real value for the peak will be displayed.

5. To reset the peak, press and hold the RESET SILENCE button until you

hear a beep, then release it.

6. To exit Peak Hold Mode and return to Leak Check Mode, press and hold

the RESET SILENCE button until you hear the GX-2012’s second beep
and the Leak Check Mode Screen displays.

Snap Log Mode, Leak Check Mode Operation

The snap logging function in Snap Log Mode allows the user to record data at
a specific time and have it saved to the data logger. The data is assigned a
snap log ID and is saved with the station ID that was selected when the data
was taken.
To enter Snap Log Mode and record snap log data:
1. Turn the GX-2012 on as described in “Turning On the GX-2012, Leak

400

PH
500
500

ppm

ppm

ppm

LEAK CK MODE

CH

4

F.S.

PEAK