GX-2012 Operator’s Manual

Display Mode • 49

Clear Data Logger Screens

CAUTION: Once you clear the data logger, you cannot retrieve any data

previously stored in the data logger.

NOTE: This item appears in Display Mode only if the DLOG DSP menu

item in Maintenance Mode is turned ON. The factory setting for
DLOG DSP is ON.

The clear data logger screens allow you to clear the data logger storage. You
can set the GX-2012 to overwrite the oldest data when the data log is full. See
“Updating the Datalog Overwrite Setting” on page 129. The bottom line of
the screen alternates between LG CLEAR, NO /DISP, and YES/ENT.

To clear the data log:
1. With the above screen displayed, press and release the POWER ENTER

button. The instrument will ask you to confirm that you want to clear the
data log and CLr LOG will be flashing. The bottom line of the screen
alternates between CONFIRM?, YES/ENT., AND NO /DISP.

2. To cancel the clear data log and return to the Clear Data Logger Screen in

Display Mode, press and release DISPLAY (ADJ).
To continue with the clear data log, press and release POWER ENTER.

CLr


LG CLEAR

LOG

CLr


CONFIRM?

LOG