GX-2012 Operator’s Manual

Using Maintenance Mode • 121

1. Use the AIRS or (SHIFT)T buttons to display the ALARM--P menu

item. Press and release the POWER ENTER button.

2. You may then select a channel. Use the AIRS or (SHIFT)T buttons to

select a channel. The combustible gas channel is shown for example.

3. Press the POWER ENTER button to display the warning setting for the

combustible channel. It will be flashing.

4. Use the AIRS or (SHIFT)T buttons to change the warning setting.
5. Press POWER ENTER to save the setting. The alarm setting for the

combustible gas channel will then be displayed.

6. Use the AIRS or (SHIFT)T buttons to change the alarm setting.

- - -


%LEL

MAINTENANCE

CH

4

ALARM--P

10


%LEL

MAINTENANCE

CH

4

WARNING

50


%LEL

MAINTENANCE

CH

4

ALARM