Beep. set, Over wrt, Updating the confirmation beep setting – RKI Instruments GX-2012 Manual User Manual
Page 137: Updating the datalog overwrite setting
![background image](/manuals/571371/137/background.png)
GX-2012 Operator’s Manual
Using Maintenance Mode • 129
7. To exit the BUMP--SET menu, use the (SHIFT)T button to scroll to
ESCAPE and press and release the POWER ENTER button.
Updating the Confirmation Beep Setting
With BEEP. SET set to ON, the GX-2012 beeps once every 5 minutes to
verify that it is operating.
With BEEP. SET set to OFF (factory setting), the GX-2012 does not sound a
confirmation beep.
1. Use the AIRS or (SHIFT)T buttons to display the BEEP. SET menu
item. Press and release the POWER ENTER button. The screen will
indicate the current setting.
2. Use the AIRS or (SHIFT)T button to display the desired setting.
3. Press and release the POWER ENTER button to save the setting and
return to the main menu.
Updating the Datalog Overwrite Setting
With OVER WRT set to ON (factory setting), the GX-2012 writes over the
oldest data with new data when the data logger memory is full.
With OVER WRT set to OFF, the GX-2012 stops saving data to the data
logger when the data logger memory is full.
1. Use the AIRS or (SHIFT)T buttons to display the OVER WRT menu
item. Press and release the POWER ENTER button. The screen will
indicate the current setting.
2. Use the AIRS or (SHIFT)T button to display the desired setting.
On
MAINTENANCE
BEEP. SET
On
MAINTENANCE
OVER WRT