GX-2012 Operator’s Manual

Using Maintenance Mode • 133

Updating the Bump Test Reminder Setting

With BP RMNDR set to ON, the GX-2012 will give an indication at start up
if it is due for bump testing. The type of indication will depend on the BP
EXPRD
setting (see below).
With BP RMNDR set to OFF (factory setting), the GX-2012 will give no
indication at start up if it is due for bump testing.

NOTE: If the BUMP DSP menu item is set to OFF, this menu item does not

appear.

1. Use the AIRS or (SHIFT)T buttons to display the BP RMNDR menu

item. Press and release the POWER ENTER button. The screen will
indicate the current setting.

2. Use the AIRS or (SHIFT)T button to display the desired setting.
3. Press and release the POWER ENTER button to save the setting and

return to the main menu.

Updating the Bump Test Expired Action Setting

This item defines what indication is given during start up when a bump test is
due and BP RMNDR is set to ON.
With BP EXPRD set to CONFIRM (factory setting), the GX-2012 will give
an indication at start up if a bump test is due and require the user to press the
RESET SILENCE button to continue.
With BP EXPRD set to NOT USE, if the unit is due for a bump test the GX-
2012 will give an indication at start up that the bump test is due and the unit
cannot be used until it is bump tested. With this setting, it is necessary to
bump test the instrument from the Maintenance Mode if a bump test is not
performed before it is due. See “Performing a Bump Test” on page 111 for
instructions.
With BP EXPRD set to NO EFFECT, no confirmation during startup or
bump test is required to use the GX-2012.

On


MAINTENANCE

BP RMNDR