Poff dsp, Bump--set, Gas time – RKI Instruments GX-2012 Manual User Manual

Page 135: Updating the bump test parameters

background image

GX-2012 Operator’s Manual

Using Maintenance Mode • 127

2. Press and release the POWER ENTER button. The screen will indicate

the current setting.

3. Use the AIRS or (SHIFT)T button to display the desired setting.
4. Press and release the POWER ENTER button to save the setting and

return to the main menu.

Updating the Bump Test Parameters

The BUMP--SET menu item is used to adjust the bump test parameters.

NOTE: If the BUMP DSP menu item is set to OFF, this menu item does not

appear.

1. Use the AIRS or (SHIFT)T buttons to display the BUMP. SET menu

item.

2. Press and release the POWER ENTER button. The values displayed are

for GAS TIME, CHECK (bump test tolerance), CAL TIME, and AUTO
CAL. The cursor on the left of the screen shows which setting is
associated with the parameter shown at the bottom of the screen. Use the
AIRS or (SHIFT)T buttons to scroll up or down along the list.

OFF


MAINTENANCE

POFF DSP


MAINTENANCE

BUMP--SET

30
30
90
On

MAINTENANCE

GAS TIME