GX-2012 Operator’s Manual
Using Maintenance Mode • 127
2. Press and release the POWER ENTER button. The screen will indicate
the current setting.
3. Use the AIRS or (SHIFT)T button to display the desired setting.
4. Press and release the POWER ENTER button to save the setting and
return to the main menu.
Updating the Bump Test Parameters
The BUMP--SET menu item is used to adjust the bump test parameters.
NOTE: If the BUMP DSP menu item is set to OFF, this menu item does not
appear.
1. Use the AIRS or (SHIFT)T buttons to display the BUMP. SET menu
item.
2. Press and release the POWER ENTER button. The values displayed are
for GAS TIME, CHECK (bump test tolerance), CAL TIME, and AUTO
CAL. The cursor on the left of the screen shows which setting is
associated with the parameter shown at the bottom of the screen. Use the
AIRS or (SHIFT)T buttons to scroll up or down along the list.
OFF
MAINTENANCE
POFF DSP
MAINTENANCE
BUMP--SET
30
30
90
On
MAINTENANCE
GAS TIME