S id sel, Saved – RKI Instruments GX-2012 Manual User Manual
Page 43

GX-2012 Operator’s Manual
Measuring Mode, Normal Operation • 35
day, and time of the snap log, and that you should press the POWER
ENTER button to save a set of snap log data. The fourth screen displays
the current gas readings, and that you should press the DISPLAY (ADJ)
button to exit Snap Log Mode without saving a set of snap log data.
2. You can change the Station ID to be used with the snap log by pressing
the (SHIFT)T button and then pressing the DISPLAY (ADJ) button. The
Station ID Select Screen will appear and the current Station ID will be
flashing.
3. Use the (SHIFT)T and AIRS buttons to scroll to the desired station ID,
then press and release the POWER ENTER button to return to the Snap
Logging Screen sequence.
To return to the Snap Logging Screen sequence without changing the
station ID, press and release RESET SILENCE.
4. To take a snap log of the current gas readings, press and release the
POWER ENTER button. The unit will display SAVED along the bottom
of the screen before returning to the Snap Logging Screen sequence.
To exit Snap Log Mode without taking a snap log or when you are
finished recording snap logs, press and release the DISPLAY (ADJ)
button. The unit will immediately return to the Normal Operation Screen.
5. The data recorded in Snap Log Mode can be viewed in Display Mode.
See “Snap Logging Screen” on page 53 for more information.
S Id
SEL
001