GX-2012 Operator’s Manual

Bump Testing • 173

gas flow, clamp down the hose clamp and remove the regulator or
dispensing valve from the cylinder.

6. Disconnect the tubing from the regulator or dispensing valve.
7. Install the hose and probe to the GX-2012.
8. Use the AIRS or (SHIFT)T buttons to display the BUMP menu item.

9. Press and release the POWER ENTER button to display the bump test

gas values.

10. Use the AIRS or (SHIFT)T buttons to see the %volume combustible

gas bump test value.

11. A bump test may be performed on the standard 4 sensors or on the

%volume combustible sensor. Use the AIRS or (SHIFT)T buttons to
display the screen for the sensor(s) you wish to bump test.

12. Before proceeding, confirm that the bump test gas value(s) are the same

as the concentration(s) in the calibration cylinder. If they are not, adjust
the bump test gas value(s) by entering the AUTO CAL menu item,
changing the values there, and reentering the BUMP menu.


MAINTENANCE

BUMP

50

12.0
50
25.0

%LEL

%

ppm

ppm

MAINTENANCE

CH

4

O

2

BUMP 30

CO

H

2

S

100


vol%

MAINTENANCE

CH

4

BUMP 30