GX-2012 Operator’s Manual

Calibration Mode • 59

Calibrating the GX-2012

From Measuring Mode, enter Calibration Mode by pressing and holding the
(SHIFT)T button, then pressing the DISPLAY(ADJ) button and releasing
both. If the unit prompts you for the password, enter it by using the AIRS
and (SHIFT)T buttons to select each password number and then pressing
and releasing POWER ENTER to confirm it and move on to the next number.
To go back a number, press and release the RESET SILENCE button. The
factory set password is the same as the factory set password for Maintenance
Mode: 2102. You can change the password by entering the PASSWORD
menu item in Maintenance Mode. See “Turning the Password Function On or
Off” on page 141. If you enter an incorrect password, an error screen will
display.

To return to normal operation, press and release the RESET SILENCE
button. You may then try to reenter Calibration Mode.
MAINTENANCE will appear along the top of the screen once you have
entered Calibration Mode. The first item in the calibration menu, AIR CAL,
displays.

NOTE: The following screens illustrate a four-gas GX-2012 with both a

%LEL and a %volume combustible gas sensor as examples only.
Your GX-2012 may display slightly different screens.

The GX-2012’s calibration menu includes two methods of calibration:
auto calibration and single calibration.
Auto Calibration: This method allows you to calibrate the combustible

gas (%LEL sensor only), OXY, H

2

S, and CO sensors simultaneously. It is

Err


PASSWORD


MAINTENANCE

AIR CAL