144 • Troubleshooting

GX-2012 Operator’s Manual

• Display indicates

“FAIL SYSTEM”

during startup

along with an error

code.

• A circuit fault has

occurred on one

of the PCB’s.

1. If the error code is 031,

indicating a FLASH memory

failure, press and release the

RESET SILENCE button to

continue to measuring mode

keeping in mind that there will

be no datalogging function.

2. If the error code is anything

else, the instrument cannot be

used.

3. Contact RKI Instruments, Inc.

for further instruction if either

1 or 2 occur.

• Display indicates

“FAIL CLOCK”

during startup.

• An internal clock

failure has

occurred

because the date

in the unit’s

memory is not a

valid date.

1. Enter normal operation by

pressing and releasing the

RESET SILENCE button to

use the unit temporarily.

2. Attempt to set the date in the

DATE menu item in

Maintenance Mode.

3. If you cannot set the date,

contact RKI Instruments, Inc.

for further instruction.

• Heart symbol at

the top of the

screen becomes

steadily on or

disappears

• A microprocessor

error has

occurred.

• Contact RKI Instruments, Inc.

for further instruction.

• The LCD shows

abnormally high or

low readings but

other gas

detection

instruments do

not.

• The LEL or CO

filters may need

to be replaced

• The GX-2012

may need to be

recalibrated.

• The sensor for

the affected

channel(s) may

need

replacement.

1. If the abnormal reading is on

the LEL channel, replace the

LEL filter. If the abnormal

reading is on the CO channel,

replace the charcoal filter.

2. Recalibrate the unit.

3. If the difficulties continue,

replace the sensor for the

affected channel(s) and

calibrate the affected

channel(s).

4. If the difficulties continue,

contact RKI Instruments for

further instruction.

Table 8: Troubleshooting the GX-2012

Symptoms

Probable

Causes

Recommended Action