Lg intvl, Updating the alarm point settings – RKI Instruments GX-2012 Manual User Manual
Page 128
120 • Using Maintenance Mode
GX-2012 Operator’s Manual
2. Press and release the POWER ENTER button. The screen will indicate
the current setting.
3. Use the AIRS or (SHIFT)T button to display the desired setting.
4. Press and release the POWER ENTER button to save the setting and
return to the main menu.
Updating the Alarm Point Settings
The ALARM--P menu item allows you to update one or more warning,
alarm, STEL, and TWA (for CO and H
2
S) settings (the reading at which the
GX-2012 recognizes these conditions).
For the combustible gas channel, it is not possible to adjust either the
Warning or Alarm settings to be higher than 60 %LEL. In addition, the
Warning setting cannot be adjusted higher than the Alarm setting for all
channels.
NOTE: It is not necessary to adjust every setting for every channel when
you enter this menu item. All settings are shown here for example.
Table 7: Factory Set Alarm Points
Channel
Warning
Alarm
STEL
TWA
Combustible Gas
10% LEL
50% LEL
N/A
N/A
Oxygen
19.5% decreasing
23.5% increasing
N/A
N/A
CO
25 ppm
50 ppm
200 ppm
25 ppm
H
2
S
5 ppm
30 ppm
5 ppm
1 ppm
300
MAINTENANCE
LG INTVL