120 • Using Maintenance Mode

GX-2012 Operator’s Manual

2. Press and release the POWER ENTER button. The screen will indicate

the current setting.

3. Use the AIRS or (SHIFT)T button to display the desired setting.
4. Press and release the POWER ENTER button to save the setting and

return to the main menu.

Updating the Alarm Point Settings

The ALARM--P menu item allows you to update one or more warning,
alarm, STEL, and TWA (for CO and H

2

S) settings (the reading at which the

GX-2012 recognizes these conditions).

For the combustible gas channel, it is not possible to adjust either the
Warning or Alarm settings to be higher than 60 %LEL. In addition, the
Warning setting cannot be adjusted higher than the Alarm setting for all
channels.

NOTE: It is not necessary to adjust every setting for every channel when

you enter this menu item. All settings are shown here for example.

Table 7: Factory Set Alarm Points

Channel

Warning

Alarm

STEL

TWA

Combustible Gas

10% LEL

50% LEL

N/A

N/A

Oxygen

19.5% decreasing

23.5% increasing

N/A

N/A

CO

25 ppm

50 ppm

200 ppm

25 ppm

H

2

S

5 ppm

30 ppm

5 ppm

1 ppm

300