GX-2012 Operator’s Manual

Overview • 143

Chapter 7: GX-2012 Maintenance

Overview

This chapter describes troubleshooting procedures for the GX-2012. It also
includes procedures for recharging the battery pack (rechargeable version),
replacing the alkaline batteries (alkaline version), and replacing various
consumable parts.

WARNING: RKI Instruments, Inc. recommends that service, calibration,

and repair of RKI instruments be performed by personnel
properly trained for this work. Replacing sensors and other
parts with original equipment does not affect the intrinsic
safety of the instrument.

Troubleshooting

The troubleshooting table describes error messages, symptoms, probable
causes, and recommended action for problems you may encounter with the
GX-2012.

Table 8: Troubleshooting the GX-2012

Symptoms

Probable

Causes

Recommended Action

• The LCD is blank.

• The unit may

have been turned

off.

• The alkaline

batteries may

need to be

replaced or the

lithium ion

battery pack

recharged.

1. To turn on the unit, press and

briefly hold the POWER

ENTER button.

2. If the unit does not turn on,

replace the alkaline batteries

or recharge the lithium ion

battery pack.

3. If the difficulties continue,

contact RKI Instruments, Inc.

for further instruction.