beautypg.com

Password, Turning the password function on or off – RKI Instruments GX-2012 Manual User Manual

Page 149

background image

GX-2012 Operator’s Manual

Using Maintenance Mode • 141

Turning the Password Function On or Off

With PASSWORD set to ON, the GX-2012 prompts you for a password
when you enter Calibration Mode.
With PASSWORD set to OFF (factory setting), no password is required to
enter Calibration Mode.

NOTE: A password is always required to enter Maintenance Mode even if

PASSWORD is turned OFF. You may change the factory set
password by turning PASSWORD ON and entering a new
password when prompted for it.

1. Use the AIRS or (SHIFT)T buttons to display the PASSWORD menu

item. Press and release the POWER ENTER button. The screen indicates
the current setting.

2. Use the AIRS or (SHIFT)T button to display the desired setting.
3. If you selected OFF, press and release the POWER ENTER button to

save the setting and return to the main menu.
If you selected ON, press and release the POWER ENTER button. ON is
replaced by the four characters making up the current password.

4. Use the AIRS or (SHIFT)T button to select a number from 0 to 9.
5. Press and release the POWER ENTER button to enter the selection and

advance to the next character.

6. Repeat steps 4 and 5 to select the remaining characters. When you press

and release the POWER ENTER button to enter the last number, the

On


MAINTENANCE

PASSWORD

2102


MAINTENANCE

PASSWORD