Cl intvl, Cl rmndr, Updating the calibration interval – RKI Instruments GX-2012 Manual User Manual
Page 138: Updating the calibration reminder setting

130 • Using Maintenance Mode
GX-2012 Operator’s Manual
3. Press and release the POWER ENTER button to save the setting and
return to the main menu.
Updating the Calibration Interval
This CL INTVL setting defines the amount of time between calibrations.
The minimum setting is 1 day and the maximum setting is 365 days. The
factory setting is 90 days.
1. Use the AIRS or (SHIFT)T buttons to display the CL INTVL menu
item. Press and release the POWER ENTER button. The screen will
indicate the current setting.
2. Use the AIRS or (SHIFT)T button to display the desired setting.
Tip: Press and hold the AIRS or (SHIFT)T button to rapidly scroll
through the settings.
3. Press and release the POWER ENTER button to save the setting and
return to the main menu.
Updating the Calibration Reminder Setting
With CL RMNDR set to ON (factory setting), the GX-2012 will give an
indication at start up if it is due for calibration. The type of indication will
depend on the CL EXPRD setting (see below).
1. Use the AIRS or (SHIFT)T buttons to display the CL RMNDR menu
item. Press and release the POWER ENTER button. The screen will
indicate the current setting.
2. Use the AIRS or (SHIFT)T button to display the desired setting.
90
MAINTENANCE
CL INTVL
On
MAINTENANCE
CL RMNDR