GX-2012 Operator’s Manual
Start Up, Normal Mode • 21
•
If the unit is due for calibration and CL EXPRD is set to NOT USE,
then the following screen displays.
The GX-2012 cannot be used until a calibration has been performed
either by selecting AUTO CAL or ONE CAL in the Maintenance
Mode menu. See “Calibrating Using Auto Calibration” on page 105
or “Calibrating Using Single Calibration” on page 108 for calibration
instructions.
•
If calibration is not due or CL EXPRD is set to NO EFFECT, then
the following screen appears for a few seconds indicating when the
next calibration is due.
6. If BP RMNDR is turned on (factory setting is OFF), the screen that
appears next depends on how BP EXPRD is set in the Maintenance
Mode Menu (page 131).
•
If the unit is due for a bump test and BP EXPRD is set to CONFIRM
(factory setting), then the following screen displays.
The alarm LED’s and buzzer will pulse several times. After this,
press the RESET SILENCE button until you hear a beep to continue.