Cal sel, Confirm, Updating the calibration expired action setting – RKI Instruments GX-2012 Manual User Manual
Page 139: Updating the calibration check setting

GX-2012 Operator’s Manual
Using Maintenance Mode • 131
3. Press and release the POWER ENTER button to save the setting and
return to the main menu.
Updating the Calibration Expired Action Setting
This item defines what indication is given during start up when calibration is
due and CL RMNDR is set to ON.
With CL EXPRD set to CONFIRM (factory setting), the GX-2012 will give
an indication at start up if calibration is due and require the user to press the
RESET SILENCE button to continue.
With CL EXPRD set to NOT USE, if the unit is due for calibration the GX-
2012 will give an indication at start up that calibration is due and the unit
cannot be used until it is calibrated. With this setting, it is necessary to
calibrate the instrument from Maintenance Mode if calibration is not
performed before it is due. See “Calibrating Using Auto Calibration” on
page 105 and “Calibrating Using Single Calibration” on page 108 for
instructions.
With CL EXPRD set to NO EFFECT, no confirmation during startup or
calibration is required to use the GX-2012.
1. Use the AIRS or (SHIFT)T buttons to display the CL EXPRD menu
item. Press and release the POWER ENTER button. The screen will
indicate the current setting.
2. Use the AIRS or (SHIFT)T button to display the desired setting along
the bottom of the screen.
3. Press and release the POWER ENTER button to save the setting and
return to the main menu.
Updating the Calibration Check Setting
This item setting defines which sensors are used to update the calibration
reminder screen. C--CHECK can be set to 4-GAS or ALL. If C--CHECK is
set to 4-GAS (factory setting), the calibration dates for only the standard 4
sensors (LEL, O
2
, H
2
S, and CO) are used to determine if calibration is due. If
C--CHECK is set to ALL, the calibration dates for the standard 4 sensors
along with the %volume sensor are used to determine if a calibration is due.
CAL
SEL
MAINTENANCE
CONFIRM