beautypg.com

Air cal, Hold air, Performing a fresh air adjustment, normal mode – RKI Instruments GX-2012 Manual User Manual

Page 35

background image

GX-2012 Operator’s Manual

Start Up, Normal Mode • 27

NOTE: If the CO DISP parameter in Maintenance Mode is set to On, the

CO channel will appear in Leak Check Mode. If the CO DISP
parameter is set to OFF (factory setting), the CO channel will not
appear in Leak Check Mode. See “Chapter 5: Using the GX-2012 in
Leak Check Mode” on page 84 for more information.

You can scroll up or down through these screens using the AIRS or
(SHIFT)T buttons.

6. With the Normal Mode Select Screen displayed, press and release the

POWER ENTER button and continue with the startup sequence as
described in the previous section, “Turning On the GX-2012, Normal
Mode Only”, from step 4 on page 20 on.

NOTE: If no button is pressed for 20 seconds, the unit will proceed into

whichever mode is displayed.

7. To exit Normal Mode and return to the Mode Select Screen, press and

hold the (SHIFT)T button for 5 seconds while in the Normal Operation
Screen.

Performing a Fresh Air Adjustment, Normal Mode

Before using the GX-2012, it is recommended to set the fresh air readings for
the target gases by performing a fresh air adjustment. This will set the
combustible gas, CO, and H

2

S channels to zero and the OXY channel to

20.9%.
1. Find a fresh-air environment. This is an environment free of toxic or

combustible gases and of normal oxygen content (20.9%).

2. Turn on the unit as described above in “Turning On the GX-2012,

Normal Mode Only” or “Turning On the GX-2012 in Normal Mode With
All Modes Active”.

3. Press and hold the AIRS button. The display prompts you to hold the

AIRS button.

4. Continue to hold the AIRS button until the display prompts you to

release the AIRS button. The GX-2012 will count down from 8 on an
autoranging combustible gas channel as it sets the fresh air reading for all

AIr
CAL

HOLD AIR