Pump, Flow chamber, Sensors and filters – RKI Instruments GX-2012 Manual User Manual
Page 19: Inlet filter holder
GX-2012 Operator’s Manual
Instrument Description • 11
Pump
A diaphragm pump inside the GX-2012 draws the sample to the sensors. It
can draw sample from as far as 50 feet from the GX-2012. The pump is not
user serviceable.
CAUTION: Sample hose lengths of more than 50 feet are not recommended
for the GX-2012 because of flow rate reduction.
Flow Chamber
The flow chamber is on the back of the GX-2012 and is held in place by three
phillips screws. The flow chamber seals to the rubber sensor gasket which
seals to the sensor faces inside the GX-2012 and routes flow from the pump
to the sensors and to the exhaust port (also a part of the flow chamber).
Sensors and Filters
The sensors are located underneath the flow chamber and are only accessible
if you remove the flow chamber and the sensor gasket.
An H
2
S removal filter disk is placed into a recess in the sensor gasket over
the %LEL sensor. This filter darkens as it absorbs H
2
S and should be
replaced when it is a dark brown color. Check the condition of this filter
quarterly.
A charcoal filter is placed into a recess in the sensor gasket over the CO
sensor. The charcoal filter is black, has a woven texture, and is impregnated
with an H
2
S absorbing material. The CO sensor will respond if exposed to
H
2
S and certain hydrocarbon gases. The charcoal filter disk scrubs these
gases out of the sample to avoid false CO readings. If false or elevated CO
readings are noticed, especially in the presence of H
2
S, change the charcoal
filter.
Inlet Filter Holder
The filter holder is a clear plastic dome shaped piece on the top of the case. A
male quick connect fitting is located on the inlet filter holder. This is the GX-
2012’s inlet fitting. A cotton dust filter is inside the filter holder. The filter
holder may be removed by turning it counterclockwise and pulling it away
from the case. Two flat membrane disk hydrophobic filters, a wire mesh disk,
and a rubber filter retaining gasket are held in place by the filter holder and
are located in the bottom of the case chamber where the filter holder is
installed.