GX-2012 Operator’s Manual

Using Maintenance Mode • 107

sensors.

15. After the last calibration gas value is set, use the AIRS button to scroll to

ESCAPE and press POWER ENTER to return to the calibration gas
values screen.

NOTE: The RKI Four Gas Cylinder typically contains 12% O

2

by volume.

Be sure to set the “OXY” reading to agree with the concentration
listed on the cylinder’s label, not zero.

16. Use the AIRS and (SHIFT)T buttons to select which sensor(s) you want

to calibrate, the standard 4 sensors or the %volume combustible gas
sensor.
The following instructions illustrate calibrating the standard 4 sensor
first, then the %volume combustible gas sensor.

17. With the calibration values screen displayed, press the POWER ENTER

button. AUTO CAL begins to flash and the current gas readings are
displayed.

18. Connect the tubing from the demand flow regulator to the rigid tube on

the probe. Allow the gas to flow for one minute.