124 • Using Maintenance Mode

GX-2012 Operator’s Manual

14. Use the AIRS or (SHIFT)T buttons to select the H

2

S channel and press

POWER ENTER. Adjust the values for the warning and alarm settings
using the procedure in step 4 through step 7. When you press POWER
ENTER after selecting the alarm value, you will then be prompted to
select the STEL value as you were with the CO channel.

15. Use the AIRS or (SHIFT)T buttons to change the setting and press

POWER ENTER when finished. You will then be asked to enter a TWA
value.

16. Use the AIRS or (SHIFT)T buttons to change the setting and press

POWER ENTER when finished to return to the Alarm Points Screen.

17. When you are finished making adjustments, use the AIRS or (SHIFT)T

buttons to select ESCAPE and press POWER ENTER. You will be
returned to the ALARM--P screen in the main menu.

Turning the Clear Data Logger Screen On or Off

The DLOG DSP menu item allows you to turn the Clear Data Logger Screen
on or off. When set to ON (factory setting), the Clear Data Logger Screen
appears in Display Mode and allows the user to clear any logged data. When



5.0

ppm

MAINTENANCE

STEL

H

2

S




1.0

ppm

MAINTENANCE

TWA

H

2

S




- - -

ppm

MAINTENANCE

ALARM--P

H

2

S