GX-2012 Operator’s Manual

Replacing a Sensor • 155

flow chamber.

4. Grab the exhaust tube and pull it gently away from the unit to slide the

flow chamber out of the unit.

5. Gently pull on the knob of the sensor gasket to remove it. Be sure not to

lose any of the filters that are in it. The sensors will be exposed.

6. Remove the sensor from its socket.
7. Carefully insert the replacement sensor in the correct socket.

Figure 27: Replacing a Sensor

CAUTION: When replacing a sensor, verify that the sensor is properly

aligned with its socket before inserting it into the socket. The CO
and H

2

S sensors have alignment slots which match up with

alignment tabs in the sockets. The combustible sensors each
have two non-polarized contacts which must line up with the
contacts in their socket. Forcing a sensor into its socket may
damage the sensor or the socket.

Flow

Chamber

PPM

Combustible

Sensor

O2

HC

PPM

HC

LEL

Align Tabs in CO Sensor
with Tabs in Sockets

% LEL
Combustible Sensor

% Volume
Combustible Sensor

CO Sensor

Sensor
Gasket

Knob

HC

VOL

CO

Oxygen

Sensor

Exhaust