60 • Calibration Mode

GX-2012 Operator’s Manual

designed for use with the RKI 4-gas calibration cylinder and is the
quickest and easiest method to calibrate the
GX-2012. If installed, you may also calibrate the %volume sensor
separately from the standard four.

Single Calibration: This method allows you to calibrate one sensor at a

time. Use this method if you are only calibrating one or two sensors, if
you are calibrating only the %volume combustible sensor, or if you are
not using the RKI 4-gas calibration cylinder.

Calibrating with the Auto Calibration Method

This section describes calibration using the auto calibration method.
To calibrate using the single calibration method, see “Calibrating with the
Single Calibration Method” on page 63. These instructions assume you are
using a 4-gas cylinder and a %volume methane cylinder.

NOTE: If the combustible gas channel(s) are calibrated to a gas other than

methane, use an appropriate multigas cylinder or use single
calibration to calibrate the combustible channel(s).

1. You may only calibrate either the standard 4 sensors or the % volume

combustible sensor at a time. Once you are finished, you are returned to
the initial AUTO CAL screen and must begin the auto calibration process
again.

2. Use the AIRS and (SHIFT)T buttons to display the AUTO CAL menu

item, press and release the POWER ENTER button to display the
calibration values screen. If you want to exit back to the main menu
without completing a calibration, press and release the DISPLAY (ADJ)
button.